Skip to main content
. Author manuscript; available in PMC: 2015 Jan 15.
Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2013 Dec 3;0:225–232. doi: 10.1016/j.jchromb.2013.11.054

Table 3. K2 and K3EDTA Plasma Equivalence and Dilution Accuracy.

DTG in Plasma Matrix
EDTA Equivalence K2EDTA K3EDTA

Low Mid High Low Mid High

15 450 9000 theoretical conc. 15 450 9000
14.10 412.67 8676.90 mean 15.83 460.87 9236.90
1.02 22.75 59.93 SD 0.31 20.72 149.29
7.2 5.5 0.7 %CV 1.9 4.5 1.6
-6.0 -8.3 -3.6 %dev 5.6 2.4 2.6
-9.2 -3.5 -3.3 %diff* 9.0 3.5 3.3

Dilutions of Ultra High QC K2EDTA K3EDTA

1/5 dil 1/10 dil 1/20 dil 1/5 dil 1/10 dil 1/20 dil

6000 3000 1500 theoretical conc. 6000 3000 1500
5477.57 3004.63 1459.60 mean 5615.77 2863.67 1457.73
173.29 24.99 17.13 SD 184.31 58.19 45.08
3.2 0.8 1.2 %CV 3.3 2.0 3.1
-8.7 0.2 -2.7 %dev -6.4 -4.5 -2.8
*

%difference as compared to matrix matched QCs, determined by the following equation: (mean unmatched – mean matched)/[((mean unmatched + mean matched)/2)*100].