Table 3. Comparison to other stabilization and registration methods.

Method Optical stabilization (arcmin) Digital registration (arcmin) Description
Rochester AOSLO (this report) 0.34–0.53 0.04–0.05 Optical stabilization with digital registration
Berkeley AOSLO [29] N/A 0.15 Digital image registration only
PSI-IU WF-SLO [34] 3–4 N/A Utilized optic disk reflectometer for optical stabilization
Optical lever [19,20] 0.2–0.38 N/A Direct optical coupling via rigid contact lens
Dual Purkinje (DPI) eye tracker with optical deflector [21,22] 1 N/A Measures displacements of Purkinje reflexes from cornea and lens; coupled to optical deflector
EyeRis™ [23] 1 N/A DPI with gaze contingent display