. 2014 Nov 4;14:127. doi: 10.1186/1471-2415-14-127

Table 2.

Comparison of intraocular pressure (IOP) measured by goldmann applanation tonometry (GAT) and a tono-pen AVIA tonometer (Reichert technologies Inc., Depew, NY, USA) in a sitting position

Mean ± SD Range p-value
IOPGAT (mmHg) 12.81 ± 2.11 6 ~ 18 0.673*
IOPAVIA (mmHg) 12.92 ± 1.71 6 ~ 19
Correlation
  Pearson’s r 0.788 < 0.001
  Linear regression r 0.752 < 0.001
Difference (mmHg) 0.15 ± 1.09 -2 ~ 2
  Linear regression r 0.052 0.644

SD Standard deviation.

*Calculated by student's paired t-test.

IOPGAT minus IOPAVIA.

Calculated by Altman-Bland's method.