Table 3.

The precision and accuracy of dibudipine determination in rat plasma

Sample concentration (ng/mL) 10(LLOQ) a 100(low) 500(Medium) 1000(High)
Intra-day(n=5)
Mean ± S.D. 9.2 ± 0.7 91.4 ± 4.1 450.0 ± 18.7 940.0 ± 41.8
CV (%) b 8.1 4.5 4.15 4.4
Error (%) c -7.8 -7.6 -10.0 -6.0
Inter-day(n=5)
Mean ± S.D. 9.1 ± 0.3 90.8 ± 2.9 502.0 ± 37.0 920.0 ± 25.5
CV (%) 3.8 3.2 7.3 2.7
Error (%) -9.0 -9.2 0.4 -8.0

The measured concentration minus the expected concentration to the expected concentration × 100.

a

LLOQ: lower limit of quantification

b

Precision (at each concentration) was expressed as CV% = (SD/ mean measured concentration) × 100

c

Accuracy (at each concentration) was expressed as the Error%, which was calculated by dividing