Table 2.

Recommendations gathered from study for pattern recognition based myoelectric control.

Property Recommendation
Interelectrode Distance Up to 4cm, as permitted by residual limb size
Electrode Size 1cm2
Electrode Configuration Combination of both longitudinal and transverse oriented channels
Number of Channels 4–6 for transradial
EMG features Time Domain and Autoregressive
Classification Method Linear Discriminant Analysis