Skip to main content
. 2014 Nov 10;41(12):121702. doi: 10.1118/1.4899177

TABLE I.

Local percent dose differences between the multiple ion chamber phantom measurements and the planning system calculations at different locations within the 24 plans. Measurements which were greater than ±3% different from the tps planned dose are in italics, those that differ by more than 4% are in bold.

Percent differences between measured and calculated doses
Plan 1 2 3 4 5 6 7 8 9 10 Acceptable (Y/N)
GI 1 4.2% 6.0% 5.9% 5.7% 6.1% 3.5% 3.1% 5.4% 0.7% N
GI 2 5.1% 6.6% 4.8% 1.2% 2.0% 6.2% 5.4% 5.2% 1.4% 2.3% N
GI 3 −0.7% 2.0% 1.2% 5.5% 5.1% −1.2% N
GI 4 −0.4% −0.1% 0.8% 0.9% 0.7% 1.0% 1.3% 0.8% 1.0% Y
GI 5 −2.7% −1.2% −0.4% 3.2% −1.0% −2.1% −1.8% −0.7% Y
GU −2.8% −2.1% −1.6% −2.8% −2.8% −0.8% 3.3% Y
Gyn 1 −2.1% −1.8% −2.7% −0.8% −1.2% −1.9% 0.5% −0.8% −1.3% Y
Gyn 2 2.9% 1.6% 3.1% 1.7% 2.7% 2.7% 1.5% 7.4% N
HN 1 3.3% −1.7% −2.0% 3.4% −2.5% −1.4% −1.5% Y
HN 2 0.4% −1.1% −2.0% −0.2% 0.7% 1.3% −0.2% −1.2% Y
HN 3 −0.7% −0.8% −0.9% −0.9% 0.3% −2.4% Y
HN 4 −1.7% −2.7% 5.1% −1.0% −2.9% 0.0% 5.5% N
HN 5 −2.6% −0.6% −2.1% −2.1% −1.4% 3.6% −0.1% Y
Lung stereo −1.7% 4.7% −1.7% 4.9% N
Mantle −2.7% −0.9% 7.3% 5.5% 6.2% N
Meso 1 4.3% 5.3% 6.3% 4.0% 1.6% 5.3% 2.4% 5.3% 4.8% 4.7% N
Meso 2 7.4% 4.8% 5.2% 1.6% 5.4% 4.5% 3.2% 2.3% N
Meso 3 3.8% 3.6% 1.7% 3.1% 4.1% 2.0% 4.3% 2.2% N
Meso 4 6.8% 4.6% 3.4% 2.8% 1.3% 7.5% 4.0% 5.7% N
Rib stereo −2.1% 4.4% 4.9% 3.4% N
Spine stereo 6.8% −2.3% 4.8% 5.0% N
Thoracic 1 5.2% 5.2% −0.2% 6.3% 5.4% 8.0% 4.4% N
Thoracic 2 3.5% −2.9% −2.6% 3.5% 4.9% −1.6% −1.3% 3.4% 0.1% N
Thoracic 3 −2.1% −1.4% −1.2% −2.5% 3.1% −1.2% −1.8% −0.3% 3.3% −0.4% Y