Table 1.
Primary and secondary prevention, ALL ICD types | ||||||
---|---|---|---|---|---|---|
Clinical setting | VT zone | VF zone | Detection duration | Discrimination | ATP for VT zone | ATP for VF zone |
No structural heart disease, No VT substrate No fast SVTs |
≥200 bpm | 9–12 s (30–40 intervals) |
None, apart Medtronic and NayaMed |
During the charge or before charging 8 pulses, 85–88% RR interval |
||
No structural heart disease, No VT substrate Known fast SVTs |
200–250 bpm | ≥250 bpm Setting as previous for Medtronic & NayaMed |
9–12 s (30–40 intervals) |
VT zone up to 230 bpm for Medtronic & NayaMed, SQICD |
Single ATP 8 pulses, 85–88% RR interval |
Debatable clinical usefulness, but possible during charge in Medtronic, NayaMed, Biotronik, St Jude Medical |
Structural Heart disease, Possible VT substrate |
170–200 bpm | ≥200 bpm | 9–12 s (30–40 intervals) VT detection>12 s |
VT zone up to 230 bpm for Medtronic & NayaMed, SQICD |
2 ATP burst*, 20 ms scan 8 pulses, 85–88% RR interval Single ATP ramp* 8 pulses 81% RR interval |
During the charge or before charging* 8 pulses, 85–88% RR interval |
Structural heart disease, VT < 170 bpm unresponsive to ablation/drugs |
Dual VT 170–200 bpm 140–170 bpm |
≥200 bpm | 9–12 s (30–40 intervals) VT detection>12 s Slow VT > 60 s |
VT zone up to 230 bpm for Medtronic & NayaMed, SQICD |
Same as above for VT * 170–200 bpm Multiple individualized * ATP schemes for slow VT |
During the charge or before charging* 8 pulses, 85–88% RR interval |
* BIV ATP in CRT-D with ischemic heart disease.