. 2014 Oct 27;41(11):111911. doi: 10.1118/1.4898098

TABLE III.

Achieved spatial resolution (lp/mm) and noise for FBP and 100% ASIR with 200 mA (100 mA s) and 50 mA (25 mA s) tube currents at half DFOV. Veo measures were similar to that achieved for 100% ASIR, but were only obtained for the STD kernel because of incompatibility with higher frequency kernels. Values are plotted in Fig. 2.

Recon Recon kernel Resolution (1p/mm) σHU Recon kernel Resolution (1p/mm) σHU
algorithm 100 mA s 25 mA s
FBP STD 8 20.95 STD 8 43.70
DETAIL 9 26.27 DETAIL 9 55.89
BONE 11 78.1 BONE 11 162.23
EDGE 13 149.85 EDGE 13 307.63
ASIR STD 8 11.69 STD 7 24.14
DETAIL 9 12.45 DETAIL 9 26.34
BONE 11 34.85 BONE 11 71.97
EDGE 13 67.2 EDGE 13 174.29