Figure 1. Accuracy of PIMA compared to BD FACS.

Figure 1

Linear regression analysis plots (right column) and Bland Altman bias plots (left column) with upper and lower 95% limits of agreement (LOA) shown as broken lines and mean bias shown as an unbroken line. A. and B. PIMA versus BD FACS both performed at an urban (reference) laboratory, by a single laboratory technician, C. and D. BD FACS versus PIMA both performed at the rural laboratory by a single laboratory technician, E. and F. BD FACS performed by laboratory technician at an urban laboratory PIMA versus PIMA performed by clinical staff at an urban clinic, G. and PIMA. BD FACS performed at by laboratory technician at a rural laboratory versus PIMA performed at two rural clinics by clinical staff.