Table 4. An evaluation of the non-instrumented nucleic acid amplification (NINA) heaters as compared to a real-time PCR thermocycler by measuring the performance of the HIV LAMP assay with a range of HIV RNA target concentrations.

NINA incubator Thermocycler
150 copies/reaction 21/21 21/21
75 copies/reaction 21/21 20/21
38 copies/reaction 18/21 16/21
19 copies/reaction 17/21 13/21
Negative control 0/21 0/21

All results were confirmed by NALF, agarose gel electrophoresis, and melt curve analysis. Table entries show the number of replicates that returned a positive result as the numerator of a fraction showing the total number of replicates run in the denominator.