Table 4. An evaluation of the non-instrumented nucleic acid amplification (NINA) heaters as compared to a real-time PCR thermocycler by measuring the performance of the HIV LAMP assay with a range of HIV RNA target concentrations.
NINA incubator | Thermocycler | |
150 copies/reaction | 21/21 | 21/21 |
75 copies/reaction | 21/21 | 20/21 |
38 copies/reaction | 18/21 | 16/21 |
19 copies/reaction | 17/21 | 13/21 |
Negative control | 0/21 | 0/21 |
All results were confirmed by NALF, agarose gel electrophoresis, and melt curve analysis. Table entries show the number of replicates that returned a positive result as the numerator of a fraction showing the total number of replicates run in the denominator.