Table III.

In Vitro Aerosolization Performance with Capsules40.6 in Aerolizer at 90 L/min. Emitted Fraction (EF), Fine Particle Fraction (FPF), Respiratory Fraction (RF), Mass Median Aerodynamic Diameter (MMAD), and Geometric Standard Deviation (GSD) Obtained from Formulations Containing SBS Blended with Granulated Lactose Carriers GL 212–250 μm, GL 250–300 μm, GL 300–425 μm, GL 425–600 μm, GL 600–850 μm, and GL 850–1,000 μm (mean ± SD, n = 3)

GL 212–250 μm GL 250–300 μm GL 300–425 μm GL 425–600 μm GL 600–850 μm GL 850–1,000 μm
EF (%) 83.2 (3.5) 83.6 (5.5) 81.1 (8.0) 76.7 (2.8) 73.2 (7.3) 69.6*(11.0)
FPF (%) 36.5 (3.1) 33.8 (5.2) 35.3 (7.0) 32.8 (3.7) 36.8 (6.0) 41.6* (5.6)
RF (%) 30.4 (3.7) 28.5 (6.6) 29.0 (8.3) 25.1 (2.7) 27.1 (6.5) 29.3 (8.4)
MMAD (μm) 1.69 (0.03) 1.68 (0.11) 1.65 (0.04) 1.67 (0.03) 1.58 (0.03) 1.66 (0.01)
GSD 1.8 (0.1) 1.8 (0.2) 1.8 (0.2) 1.8 (0.1) 1.9 (0.0) 2.0 (0.2)