Table 2.

Gradient Delay Volume Measured by Each Approach under Different Experimental Conditions on Instrument A

Gradient Length (min) Flow Rate (mL/min) Stationary Phase Batch Column Dimensions Measured by MYG Measured by Modified Conventional Procedure
Mixing Volume (mL) Total Delay Volume (mL) Mixing Volume (mL) Total Delay Volume (mL)
5 0.4 2 2.1x100 mm 0.158 0.364 0.066 0.341
10 0.4 2 2.1x100 mm 0.095 0.356 0.070 0.360
10 0.1 2 2.1x100 mm 0.161 0.36 0.073 0.295
10 0.2 2 2.1x100 mm 0.105 0.355 0.086 0.322
10 0.4 1 2.1x150 mm 0.045 0.338
10 0.4 1 2.1x100 mm 0.087 0.359
10 0.4 1 2.1x50 mm 0.08 0.373
10 0.4 1 3.0x100 mm 0.068 0.344
10 0.4 3 2.1x100 mm 0.055 0.331
10 0.4 4 2.1x100 mm 0.092 0.366
10 0.4 5 2.1x100 mm 0.095 0.368
10 0.4 6 2.1x100 mm 0.057 0.345

Average: 0.092±0.037 0.355±0.013 0.074±0.009 0.329±0.028