. 2014 May 5;43(2):235–243. doi: 10.1111/vcp.12150

Table 2.

Limit of quantification (LoQ) determined in three canine serum pools diluted to low C-reactive protein (CRP) concentrations. Samples were analyzed 36 times on 3 different days with a new automated canine-specific immunoturbidimetric CRP method. The total error (TE) was calculated (TE = bias + 2 standard deviations [SD]). LoQ was 6.8 mg/L, as TE exceeded the set quality goal (29.6%) for lower concentrations.

Expected CRP Concentration (mg/L) Mean Observed CRP Concentration (mg/L) SD (mg/L) TE (%) TE < TEa
6.8 7.1 0.39 16 Yes
4.9 5.9 0.35 38 No
2.9 4.8 0.39 90 No

TEa indicates allowable total error.