Interference: the observed difference in C-reactive protein (CRP) concentration between a canine test pool with interfering substance and control pool, when measured with a new automated canine-specific immunoturbidimetric CRP method. The maximum allowable change in CRP concentration caused by interference was set to ± 10%.

Interferent CRPcontrol pool (mg/L) CRPtest pool (mg/L) dobs [95% CI] (mg/L) Acceptance Limits* of Interference (mg/L)
Hemoglobin 10 g/L 29.0 29.3 0.33 [−1.28; 1.95] [−2.9; 2.9]
111.5 111.5 −0.03 [−2.39; 2.32] [−11.2; 11.2]
Triglycerides 10 g/L 30.3 31.3 1.00 [−0.62; 2.62] [−3.0; 3.0]
119.5 119.8 0.27 [−2.08; 2.62] [−12.0; 12.0]

dobs (95 CI) indicates observed difference (95% confidence interval).

*

Acceptance limits = ± 0.1 × CRPcontrol pool.