Precision of the RSA measurements a

Axis Translation, mm Rotation, °
Transverse (X) 0.12 0.23
Longitudinal (Y) 0.08 0.30
Sagittal (Z) 0.12 0.09

a The precision is based on 50 double examinations of the patients in the study. The value given represents the smallest migration that is considered statistically significant and is based on 2 standard deviations of the error obtained. This therefore corresponds to the 95% confidence limit.