Skip to main content
. 2013 Dec 7;4(4):234–241. doi: 10.1016/j.jpha.2013.11.004

Table 2.

Linearity, intra- and interday accuracy and precision of analytical method for NAE measurement from brain matrix.

Analyte R2 Nominal concentration (ng/mL) Intraday 1 (n=3)
Intraday 2 (n=3)
Interday (n=8)
Mean (ng/mL) Accuracy (%) CV (%) Mean (ng/mL) Accuracy (%) CV (%) Mean (ng/mL) Accuracy (%) CV (%)
AEA 0.999 7.5 8.6 115.0 9.4 7.9 105.8 1.3 8.2 109.5 7.8
35 31.7 105.7 6.2 34.2 114.0 6.1 33.2 110.8 7.5
75 68.1 90.8 7.5 72.5 96.7 9.1 68.5 91.4 8.2



OEA 0.989 15 18.2 121.5 18.8 14.8 98.4 5.5 15.9 105.8 17.6
30 30.9 102.9 5.1 25.1 83.6 2.8 28.4 94.8 13.3
90 85.8 95.4 5.3 90.6 100.7 4.2 85.9 95.4 6.1



PEA 0.999 15 14.9 99.4 11.6 13.9 93.0 3.6 14.3 95.0 8.6
30 28.5 95.0 13.2 29.1 97.1 7.2 28.6 95.8 8.7
90 94.8 105.3 5.4 80.7 89.6 4.9 88.9 98.8 8.9