Skip to main content
. 2014 Nov 19;18(6):603. doi: 10.1186/s13054-014-0603-0

Table 3.

Possible parameters that affect differences in EVVTI and EVMM using mixed-model analysis

Parameter Study arm Effect on difference of methods (%) P value
(95% limits of agreement)
Non-invasive CPAP in PREM EVVTI 40.1 (34.6/46.4) 0.022
EVMM 31.6 (26.6/37.6) 0.732
Ventilation via ET tube EVVTI 5.0 (4.3/5.9) 0.017
EVMM 4.7 (4.0/5.7) 0.069
Body weight EVVTI 9.5 (6.4/7.5) <0.001
EVMM 11.5 (9.7/13.5) <0.001
Secondary abdominal closure EVVTI 5.2 (4.4/6.1) 0.272
EVMM 6.9 (5.5/8.7) 0.002

CPAP, continuous positive airway pressure; ET, endotracheal; EVMM, electrical velocimetry versus the transthoracic echocardiography M-mode; EVVTI, electrical velocimetry versus the transthoracic echocardiography velocity time integral; PREM, preterm infants subgroup.