Treatment beam quality parameters

Electron beam R50(cm) Dmax(cm) Inline graphic (MeV) Rp(cm) Contamination (%)
Standard 4 MeV 1.2 0.6 3.2 1.6 0.3
Standard 6 MeV 2.3 1.3 5.5 2.9 0.6
4 MeV TSEI 0.95 0.4 2.9 1.3 < 0.001
6 MeV TSEI 1.2 0.4 3.2 1.6 < 0.001

Calculated using TG-70: Inline graphic.