Table 2.
Accuracy of Retention Projections Under a Range of Experimental Conditions
| Experimental Condition | Accuracy of Retention Projectionsa |
|---|---|
| 3.1.1 Metal blocks in the GC oven | |
| Without blocks in oven | ±0.43 s |
| With two Al blocks in oven | ±0.35 s |
| 3.1.2 Liner make/model/geometry | |
| Agilent, 2 mm ID, 78.5 mm long, straight splitless, quartz | ±0.49 s |
| Agilent, 4 mm ID, 78.5 mm long, gooseneck, glass | ±0.44 s |
| SGE, 4 mm ID, 78.5 mm long, straight, glass | ±0.38 s |
| Restek Siltek, 2 mm ID, 78.5 mm long, straight splitless, quartz | ±0.45 s |
| Restek Sky, 4 mm ID, 78.5 mm long, straight, glass, packed w/deactivated glass wool | ±0.40 s |
| Supelco, 4 mm ID, 78.5 mm long, straight, glass, packed w/deactived glass wool | ±0.40 s |
| Thermo, 5 mm ID, 105 mm long, straight FocusLiner, packed w/deactived quartz wool | ±0.42 s |
| Thermo, 5 mm ID, 105 mm long, straight, packed w/deactived quartz wool | ±0.43 s |
| 3.1.3 Injection history | |
| Before injections | ±0.51 s |
| After 7 injections of the urine sample | ±1.3 s |
| Clean liner | ±0.43 s |
| Dirty liner | ±1.7 s |
| Clean base seal | ±0.47 s |
| Dirty base seal | ±0.50 s |
| 3.1.4 Split or splitless injection | |
| Split (10:1) | ±0.42 s |
| Splitless | ±0.43 s |
| 3.1.5 Manual injection speed | |
| Fast (~0.5 sec) | ±0.42 s |
| Slow (5 sec) | ±0.44 s |
| 3.1.6 Inlet temperature | |
| 300ºC | ±0.44 s |
| 290 ºC | ±0.42 s |
| 280 ºC | ±0.44 s |
| 270 ºC | ±0.43 s |
| 250 ºC | ±0.48 s |
| 230 ºC | ±0.45 s |
| 200 ºC | ±0.44 s |
| 3.1.7 Column cut | |
| Straight | ±0.42 s |
| Angled | ±0.45 s |
| 3.1.8 Length of column in inlet | |
| 0.2 cm | ±0.44 s |
| 1.2 cm | ±0.39 s |
| 2.2 cm | ±0.37 s |
| 3.1.9 Column conditioning | |
| None | ±0.42 s |
| 3 hrs at 320ºC | ±0.45 s |
| 3.1.10 Exposure to O2 levels in industrial grade He | |
| Before | ±0.42 s |
| After 72 hours exposure to 6.2 ppm O2 at 320ºC | ±0.39 s |
| 3.1.11 Transfer line temperature | |
| 320 ºC | ±0.44 s |
| 280 ºC | ±0.53 s |
| 240 ºC | ±0.57 s |
| 200 ºC | ±0.63 s |
| 160 ºC | ±0.66 s |
| 3.1.12 MS source temperature | |
| 300 ºC | ±0.52 s |
| 220 ºC | ±0.50 s |
| 180 ºC | ±0.49 s |
| 100 ºC | ±0.52 s |
| 3.1.13 Injection amount | |
| 1 pmol | ±0.59 s |
| 10 pmol | ±0.50 s |
| 20 pmol | ±0.49 s |
| 30 pmol | ±0.50 s |
| 50 pmol | ±0.64 s |
| 100 pmol | ±1.1 s |
| 750 pmol | ±5.5 s |
| 3.1.14 Column length | |
| 30 m | ±0.44 s |
| 25 m |
|
| 20 m |
|
| 15 m |
|
| 3.1.15 Column inner diameter | |
| 0.25 mm | ±0.44 s |
| 0.32 mm |
|
| 3.1.16 Film thickness | |
| 0.25 μm | ±0.44 s |
| 1 μm |
|
Accuracy of retention projections is given as the root-mean-square error (experimental minus predicted retention time) among all 12 test compounds.
indicates the accuracy is within the threshold for a “passed” system suitability check,
indicates “passed, but questionable”, and
indicates “failed”.
Accuracy of retention times calculated from linear retention indices that were measured under the standard 26 °C/min ramp with a 30 m column of 0.25 mm ID and 0.25 μm film thickness.
After correction for the difference from the nominal inner diameter
After correction for the difference from the nominal film thickness