Published in final edited form as: J Chromatogr A. 2014 Nov 18;1373:179–189. doi: 10.1016/j.chroma.2014.11.030

Table 2.

Accuracy of Retention Projections Under a Range of Experimental Conditions

Experimental Condition Accuracy of Retention Projectionsa
3.1.1 Metal blocks in the GC oven
 Without blocks in oven ±0.43 s
 With two Al blocks in oven ±0.35 s
3.1.2 Liner make/model/geometry
 Agilent, 2 mm ID, 78.5 mm long, straight splitless, quartz ±0.49 s
 Agilent, 4 mm ID, 78.5 mm long, gooseneck, glass ±0.44 s
 SGE, 4 mm ID, 78.5 mm long, straight, glass ±0.38 s
 Restek Siltek, 2 mm ID, 78.5 mm long, straight splitless, quartz ±0.45 s
 Restek Sky, 4 mm ID, 78.5 mm long, straight, glass, packed w/deactivated glass wool ±0.40 s
 Supelco, 4 mm ID, 78.5 mm long, straight, glass, packed w/deactived glass wool ±0.40 s
 Thermo, 5 mm ID, 105 mm long, straight FocusLiner, packed w/deactived quartz wool ±0.42 s
 Thermo, 5 mm ID, 105 mm long, straight, packed w/deactived quartz wool ±0.43 s
3.1.3 Injection history
 Before injections ±0.51 s
 After 7 injections of the urine sample ±1.3 s
 Clean liner ±0.43 s
 Dirty liner ±1.7 s
 Clean base seal ±0.47 s
 Dirty base seal ±0.50 s
3.1.4 Split or splitless injection
 Split (10:1) ±0.42 s
 Splitless ±0.43 s
3.1.5 Manual injection speed
 Fast (~0.5 sec) ±0.42 s
 Slow (5 sec) ±0.44 s
3.1.6 Inlet temperature
 300ºC ±0.44 s
 290 ºC ±0.42 s
 280 ºC ±0.44 s
 270 ºC ±0.43 s
 250 ºC ±0.48 s
 230 ºC ±0.45 s
 200 ºC ±0.44 s
3.1.7 Column cut
 Straight ±0.42 s
 Angled ±0.45 s
3.1.8 Length of column in inlet
 0.2 cm ±0.44 s
 1.2 cm ±0.39 s
 2.2 cm ±0.37 s
3.1.9 Column conditioning
 None ±0.42 s
 3 hrs at 320ºC ±0.45 s
3.1.10 Exposure to O2 levels in industrial grade He
 Before ±0.42 s
 After 72 hours exposure to 6.2 ppm O2 at 320ºC ±0.39 s
3.1.11 Transfer line temperature
 320 ºC ±0.44 s
 280 ºC ±0.53 s
 240 ºC ±0.57 s
 200 ºC ±0.63 s
 160 ºC ±0.66 s
3.1.12 MS source temperature
 300 ºC ±0.52 s
 220 ºC ±0.50 s
 180 ºC ±0.49 s
 100 ºC ±0.52 s
3.1.13 Injection amount
 1 pmol ±0.59 s
 10 pmol ±0.50 s
 20 pmol ±0.49 s
 30 pmol ±0.50 s
 50 pmol ±0.64 s
 100 pmol ±1.1 s
 750 pmol ±5.5 s
3.1.14 Column length
 30 m ±0.44 s
 25 m graphic file with name nihms645480t1.jpg
 20 m graphic file with name nihms645480t2.jpg
 15 m graphic file with name nihms645480t3.jpg
3.1.15 Column inner diameter
 0.25 mm ±0.44 s
 0.32 mm graphic file with name nihms645480t4.jpg
3.1.16 Film thickness
 0.25 μm ±0.44 s
 1 μm graphic file with name nihms645480t5.jpg
a

Accuracy of retention projections is given as the root-mean-square error (experimental minus predicted retention time) among all 12 test compounds. Inline graphic indicates the accuracy is within the threshold for a “passed” system suitability check, Inline graphic indicates “passed, but questionable”, and Inline graphic indicates “failed”.

b

Accuracy of retention times calculated from linear retention indices that were measured under the standard 26 °C/min ramp with a 30 m column of 0.25 mm ID and 0.25 μm film thickness.

c

After correction for the difference from the nominal inner diameter

d

After correction for the difference from the nominal film thickness