MS instrumentation |
Ion trap, Q-TOF, QqQ, hybrid, Orbitrap |
TOF, Q-TOF, QqQ, single-quad, Orbitrap |
Ionisation |
ESI, MALDI |
ESI, APCI |
Detector |
MCP, electron multiplier, Orbitrap |
MCP, electron multiplier, Orbitrap |
Polarity |
Positive |
Positive/negative |
High resolution |
Required |
Optional |
High mass accuracy (MS1) |
As high as possible |
70 ppm or better |
High scan speed |
Required |
Optional |
High sensitivity |
Required |
Required/less critical |
High dynamic range |
Required |
Required |
MSn capability |
Required |
Optional (comparison to standards) |
High mass accuracy (MSn) |
Optional |
Required |
High resolution |
Optional |
Required |
Chromatographic separation |
Required |
Required (screening) |
Column chemistry |
RP (HILIC) |
RP, HILIC, others |
Nano-flow |
Required |
Normal flow preferred |
Injection volume |
0.5–10 μL |
1–100 μL |
Long columns/gradients |
Required |
Optional |
Low inter-day variability |
Optional |
Required |
Software for analysis |
Vendor-specific, commercial or free software |
Vendor-specific and limited free software |
MSMS analysis |
Automated |
Manual |
Databases |
Available for sequenced organisms |
Incomplete |
Identification of analyte |
Required |
Optional |
Use of standards |
Optional (absolute quantitation, SWATH) |
Required |