Table 2.
Test description | Example assay name CAA antigen assay | Biological matrixa | Sample required (μL)b | Maximum sample input (μL)c |
---|---|---|---|---|
20 μL assay – antigen | SCAA20 | serum | 50 | 20 |
0·5 mL assay – antigen | SCAA500 | serum | 600 | 500 |
4 mL assay – antigen | SCAA4000 | serum | 5000 | 4000 |
20 μL assay – antigen | UCAA10 | urine | 50 | 10 |
0·5 mL assay – antigen | UCAA250 | urine | 500 | 250 |
4 mL assay – antigen | UCAA2000 | urine | 2500 | 2000 |
15 mL assay – antigen | UCAA7500 | urine | 10000 | 7500 |
20 μL assay – antigen | SalCAA15 | saliva – Salivette | 1 salivette | 15 |
0·5 mL assay – antigen | SalCAA375 | saliva – Salivette | 1 salivette | 375 |
20 μL assay – antigen | WMSalCAA10 | saliva – WMSS | 50 | 10 |
0·5 mL assay – antigen | WMSalCAA250 | saliva – WMSS | 500 | 250 |
4 mL assay – antigen | WMSalCAA2000 | saliva – WMSS | 4000 | 2000 |
Test description | Example assay name SEA antigen assay | Biological matrix | Sample required (μL)d | Sample input (μL)e |
---|---|---|---|---|
antibody | SAbSEA | serum | 1 | 0·1 through 4 |
antibody | UAbSEA | urine | 1 | 1 through 20 |
antibody | SalAbSEA | saliva – Salivette | 1 salivette | 1 through 20 |
antibody | WMSalAbSEA | saliva – WMSS | 1 | 1 through 20 |
Serum and urine collection is standardized. Saliva collection, however, is quite variable; different saliva collectors are available (Malamud et al. 2005; Corstjens and Malamud, 2008) and also saliva can be collected from different parts of the mouth. Currently assays are evaluated using either whole mouth-stimulated saliva (clarified by centrifugation) or saliva collected using Salivette collectors (SARSTEDT AG & Co.).
The required sample amount is the volume needed to ensure maximum sample input. The minimum requested sample volume is 50 μL; this allows convenient sample handling (transport and storage) and extraction of the sample with TCA.
The maximum sample input translates to the maximum volume of the biological matrix (relating to the CAA concentration) that is actually used in the assay. Lower amounts are feasible but will affect the lower limit of detection (note: high infection grades may require 10- or 100-fold dilution of the sample). Higher amounts are possible when using the 500, 4000 and 15000 concentration devices; this requires reloading of the devices which can be done multiple times. Note that for urine the actual sample input is only half that of serum; a consequence of the fact that TCA extraction of serum samples generates a large pellet (about 50% of the total volume), whereas urine commonly generates only a small pellet. The composition of saliva is quite variable and TCA extractions in general generate a larger pellet than observed for urine samples, but significantly less than the serum samples.
In the foreseen combined testing approach, supplying 50 μL for the antigen assays would also allow testing for antibodies.
For antibody testing the sample input is flexible, however the volume and dilution will affect the cut-off threshold. Cut-off levels are best determined from a group representing the same ethnicity and geography.