Table S1.

Overall number of handling errors, in the second attempted use of the ELLIPTA® DPI or BREEZHALER™, following nonverbal demonstrative instruction

Handling error Error No. ELLIPTA (n=150) n (%) [95% CI] BREEZHALER (n=150) n (%) [95% CI]
At least one error 3 (2.0) [0.4–5.7] 49 (32.7) [25.2–40.8]
Turned the inhaler upside down after loading a dose 1 0 (0.0) [0.0–2.4] 0 (0.0) [0.0–2.4]
Shook the inhaler after loading a dose 2 1 (0.7) [0.0–3.7] 0 (0.0) [0.0–2.4]
Did not close the inhaler until hearing a click 3 N/A 1 (0.7) [0.0–3.7]
Played with the side buttons 4 N/A 11 (7.3) [3.7–12.7]
Could not insert a capsule into the inhaler capsule chamber without instructions 5 N/A 0 (0.0) [0.0–2.4]
Did not hold the inhaler upright while pressing the side buttons* 6 N/A 11 (7.3) [3.7–12.7]
Took inhaling posture without releasing the side buttons fully 7 N/A 23 (15.3) [10.0–22.1]
Did not press the side buttons simultaneously until a “click” was heard before taking inhaling posture 8 N/A 10 (6.7) [3.2–11.9]
Could not open the cover without instructions 9 0 (0.0) [0.0–2.4] N/A
Covered vent holes while taking the inhaling posture 10 1 (0.7) [0.0–3.7] N/A
Did not open cover until a “click” was heard before taking inhaling posture 11 0 (0.0) [0.0–2.4] N/A
Played with the cover 12 1 (0.7) [0.0–3.7] N/A

Notes:

*

Allowable tilt ±45 degrees. Error numbers are identifiers assigned to each error for ease of reference. Errors 3–8 are applicable only to the BREEZHALER errors 9–12 are applicable only to the ELLIPTA DPI.

Abbreviations: CI, confidence interval; N/A, not applicable for this inhaler.