Skip to main content
. 2014 Sep 6;20(1):57–65. doi: 10.1016/j.rpor.2014.08.006

Table 2.

Global gamma evaluation (3 mm/3%) results of AAPM TG 119 benchmark IMRT plans.

Energy Benchmark plans Gamma passing rates (%)
Planar analysis
Volumetric analysis
2D
3D
Coronal Sagittal Coronal Sagittal
6 MV Multi-target 96.8 98.1 97.0 98.1 96.7
Prostate 95.2 95.6 96.9 95.1 97.9
Head and neck 95.4 95.8 95.6 95.3 96.1
C-shape 95.6 97.3 97.0 97.4 97.6



15 MV Multi-target 98.5 98.6 98.5 98.6 99.7
Prostate 95.8 95.6 95.4 95.5 98.7
Head and neck 95.2 96.9 95.5 97.2 98.1
C-shape 96.9 96.8 97.1 96.9 97.4