Skip to main content
. 2014 Dec 5;56(1):71. doi: 10.1186/s13028-014-0071-1

Table 2.

Relative specificity and sensitivity of in-house Schmallenberg virus (SBV) ELISA and a commercial SBV ELISA compared to virus neutralization test (VNT)

VNT In-house SBV ELISA Commercial SBV ELISA
Cattle Sheep Cattle Sheep
Pos Neg Pos Neg Pos Neg Pos Neg
Pos 64 47 1 59 5 39 9
Neg 100 100 100 100
Specificity 100% (100/100) 100% (100/100) 100% (100/100) 100% (100/100)
Sensitivity 100% (64/64) 97.9% (47/48) 92.2% (59/64) 81.2% (39/48)