Table 2. Electron Spin Resonance settings for NO, ROS and peroxynitrite Measurement.
Measurement of NO | Measurement of reactive oxygen species (ROS) and peroxynitrite | |
Centre field | 3308 G | 3388 G |
Sweep width | 80 G | 132 G |
Microwave frequency | 9.495 GHz | 9.497 GHz |
Microwave power | 50 mW | 1.25 mW |
Modulation Amplitude | 4.6 G | 1.63 G |
Modulation frequency | 86 kHz | 86 kHz |
Time constant | 81.92 ms | 40.96 ms |
Time constant | 20.48 ms | 10.24 ms |
Number of scans | 30 | 50 |