Abdominal MDCT “standard-dose” protocols utilized in the prospective trial*.

Protocol: IV Contrast Urolithiasis Supine CTC
Scanner GE HD750 GE HD 750 GE HD 750
Scan Type Helical Helical Helical
Rotation Time (sec) 0.5 0.8 0.5
Beam Collimation (mm) 40 40 40
Detector Rows 64 64 64
Pitch 0.516 0.516 0.984
Speed (mm/rot) 20.64 20.64 39.36
Detector Configuration 64 × 0.625 64 × 0.625 64 × 0.625
Slice Thickness for NI (mm) 1.25 1.25 1.25
Scan FOV Large Body Large Body Large Body
kVp 120 120 120
Smart mA Range 60-660 40-660 30-300
Noise Index 24 28 50
Reconstructions (FBP):
 DFOV 36-50 36-50 36-50
 Recon Type Standard Standard Standard
 Window W/L 400/50 400/50 400/50
 Recon Option Plus Plus Plus
 Slice Thickness (mm) 2.5 2.5 2.5
 Interval (mm) 1.5 1.5 1.5
*

The specific protocol for the accompanying low-dose series was derived by adjusting the noise index (NI)/slice thickness pairing (and mA range) to allow for a targeted 70-90% dose reduction (by DLP) relative to the “standard-dose” series.