Table 1.

Means and standard deviations of pulse oximeter and co-oximeter saturation measurements of percent SaO2, standard P50 based on each method for measuring O2 saturation, and pulse oximeter and EKG heart rate (HR) measurements in Caucasian subjects.

Pulse Oximeter A Pulse Oximeter B Co-oximeter/EKG
SaO2 (percent) 91.5±8.5 91.0±9.1 90.8±8.2
P50 (standard, mmHg) 26.0±2.2 26.9±3.0 26.2±0.6
Heart Rate (beats/minute) 125.9±38.5 125.9±38.4 127.0±37.7