Table 2.

Quantitative image analysis (signal intensity, noise, CNR, and SNR).

Parameter Dual-energy CCTA (n = 49), mean ± SD Standard CCTA (n = 53), mean ± SD P value
Signal intensity (HU)
 Left main 324.8 ± 94.2 429.4 ± 130.3 <.001
 Left anterior descending 307.3 ± 85.9 420.2 ± 120.5 <.001
 Left circumflex 301.9 ± 84.6 430.1 ± 118.6 <.001
 Right coronary 301.4 ± 90.2 436.3 ± 123.8 <.001
Noise (HU)
 Left main 25.5 ± 9.5 28.7 ± 12.7 .15
 Left anterior descending 29.4 ± 15.5 32.2 ± 18.9 .42
 Left circumflex 28.8 ± 13.0 38.0 ± 20.7 <.05
 Right coronary 26.3 ± 11.8 35.2 ± 20.9 <.05
SI and noise measurements
 Mean signal intensity (HU) 308.4 ± 84.4 429.0 ± 119.5 <.001
 Mean noise (HU) 27.5 ± 8.9 33.5 ± 14.0 <.05
 CNR 16.8 ± 5.2 16.9 ± 4.8 .95
 SNR 12.0 ± 3.9 13.8 ± 3.9 .02
 SNR (after multivariate analysis) 12.0 ± 3.9 13.8 ± 3.9 .13

CCTA, coronary CT angiography; CNR, contrast-to-noise ratio; HU, Hounsfield units; SD, standard deviation; SNR, signal-to-noise ratio.