Table 11. Image quality analysis within prospectively-ECG triggered scans in a consistently fixed range of late systole to early diastole.

Variables Female (n=88) Male (n=98) P value
Aorta noise (HU), mean ± SD 36.9±10.8 31.5±9.5 <0.001
RCA signal (HU), mean ± SD 446.5±123.3 397.4±120.2 0.007
RCA CNR, mean ± SD 14.6±4.1 15.8±4.3 0.04
RCA SNR, mean ± SD 12.1±3.3 13.0±3.3 0.07
LM signal (HU), mean ± SD 472.4±128.8 413.2±118.6 0.001
LM CNR, mean ± SD 15.3±3.3 15.9±4.3 0.26
LM SNR, mean ± SD 13.2±3.14 13.6±3.3 0.43

RCA, right coronary artery; SD, standard deviation; CNR, contrast-to-noise ratio; SNR, signal-to-noise ratio; LM: left main coronary artery.