Specificities of the Elecsys syphilis assay using lot A and specificities of comparator assays

Result by sample type Elecsys syphilis assay lot A Architect syphilis TP assay LIAISON Treponema screen Serodia-TPPA Vitros syphilis TPA Enzygnost syphilis assay Immulite 2000 syphilis screen
Routine samples
    Total no. tested 3,500 1,500 3,500 1,500 2,000
    No. confirmed positive 14 1 14 1 13
    No. indeterminate 0 0 0 0 0
    No. negative 3,486 1,499 3,486 1,499 1,987
    No. false positive 7 5 12 2 0
    Specificity (% [2-sided 95% CI])a 99.80 (99.59–99.92) 99.67 (99.22–99.89) 99.66 (99.40–99.82) 99.87 (99.52–99.98) 100.00 (99.81–100.00)
Blood donor samples
    Total no. tested 4,579 3,006 1,574 1,506 1,574
    No. confirmed positive 0 0 0 0 0
    No. indeterminate 2b 2b 0 1b 0
    No. negative 4,577 3,004 1,574 1,505 1,574
    No. false positive 3 8 0 3 0
    Specificity (% [2-sided 95% CI]) 99.93 (99.81–99.99) 99.73 (99.48–99.88) 100.00 (99.77–100.00) 99.80 (99.42–99.46) 100.00 (99.77–100.00)
Total samples
    Total no. tested 8,079 4,506 3,500 1,500 1,574 1,506 3,574
    No. confirmed positive 14 1 14 1 0 0 13
    No. indeterminate 2b 2b 0 0 0 1b 0
    No. negative 8,063 4,503 3,486 1,499 1,574 1,505 3,561
    No. false positive 10 13 12 2 0 3 0
    Specificity (% [2-sided 95% CI]) 99.88 (99.77–99.94) 99.71 (99.51–99.85) 99.66 (99.40–99.82) 99.87 (99.52–99.98) 100.00 (99.77–100.00) 99.80 (99.42–99.96) 100.00 (99.90–100.00)
a

CI, confidence interval.

b

Excluded from the specificity calculation due to indeterminate confirmation results.