TABLE 2.
Result by sample type | Elecsys syphilis assay lot A | Architect syphilis TP assay | LIAISON Treponema screen | Serodia-TPPA | Vitros syphilis TPA | Enzygnost syphilis assay | Immulite 2000 syphilis screen |
---|---|---|---|---|---|---|---|
Routine samples | |||||||
Total no. tested | 3,500 | 1,500 | 3,500 | 1,500 | 2,000 | ||
No. confirmed positive | 14 | 1 | 14 | 1 | 13 | ||
No. indeterminate | 0 | 0 | 0 | 0 | 0 | ||
No. negative | 3,486 | 1,499 | 3,486 | 1,499 | 1,987 | ||
No. false positive | 7 | 5 | 12 | 2 | 0 | ||
Specificity (% [2-sided 95% CI])a | 99.80 (99.59–99.92) | 99.67 (99.22–99.89) | 99.66 (99.40–99.82) | 99.87 (99.52–99.98) | 100.00 (99.81–100.00) | ||
Blood donor samples | |||||||
Total no. tested | 4,579 | 3,006 | 1,574 | 1,506 | 1,574 | ||
No. confirmed positive | 0 | 0 | 0 | 0 | 0 | ||
No. indeterminate | 2b | 2b | 0 | 1b | 0 | ||
No. negative | 4,577 | 3,004 | 1,574 | 1,505 | 1,574 | ||
No. false positive | 3 | 8 | 0 | 3 | 0 | ||
Specificity (% [2-sided 95% CI]) | 99.93 (99.81–99.99) | 99.73 (99.48–99.88) | 100.00 (99.77–100.00) | 99.80 (99.42–99.46) | 100.00 (99.77–100.00) | ||
Total samples | |||||||
Total no. tested | 8,079 | 4,506 | 3,500 | 1,500 | 1,574 | 1,506 | 3,574 |
No. confirmed positive | 14 | 1 | 14 | 1 | 0 | 0 | 13 |
No. indeterminate | 2b | 2b | 0 | 0 | 0 | 1b | 0 |
No. negative | 8,063 | 4,503 | 3,486 | 1,499 | 1,574 | 1,505 | 3,561 |
No. false positive | 10 | 13 | 12 | 2 | 0 | 3 | 0 |
Specificity (% [2-sided 95% CI]) | 99.88 (99.77–99.94) | 99.71 (99.51–99.85) | 99.66 (99.40–99.82) | 99.87 (99.52–99.98) | 100.00 (99.77–100.00) | 99.80 (99.42–99.96) | 100.00 (99.90–100.00) |
CI, confidence interval.
Excluded from the specificity calculation due to indeterminate confirmation results.