TABLE 3.

Specificities of the Elecsys syphilis assay using lots B and C and specificities of comparator assays

Result by sample type Elecsys syphilis assay lot B Elecsys syphilis assay lot C Architect Syphilis TP assay LIAISON Treponema screen Enzygnost syphilis assay
Routine samples
    Labor Schottdorf, Germany
        Total no. tested 1,000 1,000 1,000
        No. confirmed positive 1 1 1
        No. indeterminate 1b 1b 1b
        No. negative 998 998 998
        No. false positive 1 0 0
        Specificity (% [2-sided 95% CI])a 99.90 (99.44–100.00) 100.00 (99.63–100.00) 100.00 (99.63–100.00)
Blood donor samples
    Central Institute for Blood Transfusion and Immunology, Austria
        Total no. tested 1,049 1,049 1,049
        No. confirmed positive 0 0 0
        No. indeterminate 0 0 0
        No. negative 1,049 1,049 1,049
        No. false positive 0 0 0
        Specificity (% [2-sided 95% CI]) 100.00 (99.65–100.00) 100.00 (99.65–100.00) 100.00 (99.65–100.00)
    Ege University, Turkey
        Total no. tested 632 632 632
        No. confirmed positive 1 1 1
        No. indeterminate 0 0 0
        No. negative 631 631 631
        No. false positive 3 3 5
        Specificity (% [2-sided 95% CI]) 99.52 (98.62–99.90) 99.52 (98.62–99.90) 99.21 (98.16–99.74)
    Overall blood donor specificity (% [2-sided 95% CI]) 99.82 (99.48–99.96) 99.82 (99.48–99.96) 99.21 (98.16–99.74) 100.00 (99.65–100.00)
Overall specificity (% [2-sided 95% CI]) 99.85 (99.62–99.96) 99.89 (99.67–99.98) 99.21 (98.16–99.74) 100.00 (99.63–100.00) 100.00 (99.65–100.00)
a

CI, confidence interval.

b

Excluded from the specificity calculation due to indeterminate confirmation results.