TABLE 3.
Result by sample type | Elecsys syphilis assay lot B | Elecsys syphilis assay lot C | Architect Syphilis TP assay | LIAISON Treponema screen | Enzygnost syphilis assay |
---|---|---|---|---|---|
Routine samples | |||||
Labor Schottdorf, Germany | |||||
Total no. tested | 1,000 | 1,000 | 1,000 | ||
No. confirmed positive | 1 | 1 | 1 | ||
No. indeterminate | 1b | 1b | 1b | ||
No. negative | 998 | 998 | 998 | ||
No. false positive | 1 | 0 | 0 | ||
Specificity (% [2-sided 95% CI])a | 99.90 (99.44–100.00) | 100.00 (99.63–100.00) | 100.00 (99.63–100.00) | ||
Blood donor samples | |||||
Central Institute for Blood Transfusion and Immunology, Austria | |||||
Total no. tested | 1,049 | 1,049 | 1,049 | ||
No. confirmed positive | 0 | 0 | 0 | ||
No. indeterminate | 0 | 0 | 0 | ||
No. negative | 1,049 | 1,049 | 1,049 | ||
No. false positive | 0 | 0 | 0 | ||
Specificity (% [2-sided 95% CI]) | 100.00 (99.65–100.00) | 100.00 (99.65–100.00) | 100.00 (99.65–100.00) | ||
Ege University, Turkey | |||||
Total no. tested | 632 | 632 | 632 | ||
No. confirmed positive | 1 | 1 | 1 | ||
No. indeterminate | 0 | 0 | 0 | ||
No. negative | 631 | 631 | 631 | ||
No. false positive | 3 | 3 | 5 | ||
Specificity (% [2-sided 95% CI]) | 99.52 (98.62–99.90) | 99.52 (98.62–99.90) | 99.21 (98.16–99.74) | ||
Overall blood donor specificity (% [2-sided 95% CI]) | 99.82 (99.48–99.96) | 99.82 (99.48–99.96) | 99.21 (98.16–99.74) | 100.00 (99.65–100.00) | |
Overall specificity (% [2-sided 95% CI]) | 99.85 (99.62–99.96) | 99.89 (99.67–99.98) | 99.21 (98.16–99.74) | 100.00 (99.63–100.00) | 100.00 (99.65–100.00) |
CI, confidence interval.
Excluded from the specificity calculation due to indeterminate confirmation results.