Skip to main content
. 2014 Oct 31;14(11):20620–20644. doi: 10.3390/s141120620

Figure 1.

Figure 1.

(a) Schematic of operation of a capacitive-based membrane pressure sensor. Diaphragm deflects under pressure, changing the effective distance between two parallel plates, and thus increases the measured capacitance across the plates; (b) Circuit model of passive LC tank, commonly used for inductively coupled telemetry of capacitive-based sensors.