Table 2.

Technical parameters

1.5 T 3.0 T
SSFP cine HR-MRA SSFP cine HR-MRA
TR per line (ms) 3.1 – 3.8 2.7 – 3.3 3.0 – 3.8 2.8 – 3.5
TE (ms) 1.3 – 1.8 1.0 – 1.3 1.3 – 1.6 1.0 – 1.4
Flip angle (°) 60 – 80 12 – 30 38 – 50 14 – 20
Bandwidth (Hz/pixel) 888 – 1130 610 – 698 925 – 930 579 – 635
Field of view (mm2) 113 – 225 121 – 500 90 – 244 131 – 312
x x x x
199 – 371 196 – 500 180 – 340 280 – 500
Acquisition matrix 85 – 150 448 – 576 96 – 162 210 – 328
x x x x
192 – 256 140 – 346 192 – 208 512 – 640
Slice thickness (mm) 3.5 – 6.0 0.8 – 1.3 3.0 – 6.0 0.7 – 0.9
In-plane resolution (mm2) 1.0 – 1.7 0.8 – 1.3 0.9 – 1.6 0.6 – 1.0
x x x x
0.8 – 1.0 0.7 – 1.0 0.9 – 1.8 0.5 – 1.0
Parallel imaging factor 2 2 – 4 2 2 – 4
Number of signal averages 1 1 1 1
Specific absorption rate (W/kg) 2.5 ± 0.9 2.1 ± 0.7 2.8 ± 1.3 2.5 ± 1.3
Acquisition time 7.5 ± 2.0 (sec per slice) 20.6 ± 3.7 5.8 ± 1.7 (sec per slice) 21.2 ± 3.2

*Values are reported as minimum – maximum. Where relevant, values are reported as mean ± SD. For 1.5 T: Magnetom TIM Avanto (Siemens Medical Solutions; Malvern, PA); gradient strength = 45 mT/m, slew rate = 200 mT/m/ms. For 3.0 T: Magnetom TIM Trio (Siemens Medical Solutions; Malvern, PA), gradient strength = 40 mT/m, slew rate = 200 mT/m/ms

†Scan time for CE-MRA sequence reflects seconds/acquisition

sec seconds, TE echo time, TR repetition time