Table 1.

Diagnostic accuracy of index breathalysers compared with reference police breathalyser, using researcher interpretation of single-use breathalysers

Breathalyser Total participants Correctly positive by reference standard Correctly negative by reference standard Sensitivity (%) Specificity (%) Positive predictive value (%) Negative predictive value (%)
Alcosense Elite multiuse 205 34/38 107/167 89.5 (75.9 to 95.8) 64.1 (56.6 to 71.0) 36.2 (27.2 to 46.3) 96.4 (91.1 to 98.6)
Dräger Alco-Check single-use 108 18/19 45/89 94.7 (75.4 to 99.1) 50.6 (40.4 to 60.7) 29.0 (19.2 to 41.3) 97.8 (88.7 to 99.6)
Alcosense Single single-use 100 5/19 79/81 26.3 (11.8 to 48.8) 97.5 (91.4 to 99.3) 71.4 (35.9 to 91.8) 85.0 (76.3 to 90.8)

95% CIs are shown in parentheses.