Precision, calibration parameters, and sensitivity of the determination of GCA and BCA by HPLC.

Compound Repeatability (%) Calibration LOQd (µM) LODe (µM)
Intradaya Interdayb Regression equation r 2 Linear rangec (µM)
GCA 0.28 2.76 y = 33.52x + 0.0877 0.9998 0.50–200 0.50 0.15
BCA 0.31 3.08 y = 33.25x + 0.0971 0.9997 0.50–200 0.50 0.15

aRSD of peak area based on 8 runs of a solution of the standard over 1 day.

bRSD of peak area based on 8 runs of a solution of the standard on independent days.

cConcentration range between the limit of quantification and the upper linear limit.

dLimit of quantification: signal/noise ratio = 10.

eLimit of detection: signal/noise ratio = 3.

y = concentration (µM).

x = peak area analyte/peak area internal standard.