Published in final edited form as: Phys Med Biol. 2014 Nov 4;59(23):N211–N220. doi: 10.1088/0031-9155/59/23/N211

Table 2.

The mean optical densities for all voxels contained within spheres of specified radius representing dose responses from variously sized detectors, for one of the measured 16 mm collimator dose distributions. This data was used to fit and extrapolate the dose response down to the response of a theoretical zero-volume detector. Each data point was weighted proportionately to the square root of the number of voxels contained within that given spherical detector volume.

Radius (voxels) Mean Optical Density (scaled) Number of Voxels √(Voxels)
12 7.64 7249 85.14
11 7.76 5583 74.72
10 7.85 4193 64.75
9 7.92 3055 55.27
8 7.96 2155 46.42
7 7.98 1441 37.96
6 8.00 907 30.12
5.5 8.00 696 26.38
5 8.00 529 23.00
4.5 8.00 379 19.47
3 8.00 118 10.86
2 7.99 35 5.92
1.25 7.99 7 2.65