Skip to main content
. Author manuscript; available in PMC: 2015 Dec 7.
Published in final edited form as: Phys Med Biol. 2014 Nov 4;59(23):N211–N220. doi: 10.1088/0031-9155/59/23/N211

Table 4.

Planned and measured shot displacements and offsets for all seven experiments. The uncertainty in the measured center-to-center displacement was simulated using Monte Carlo techniques to be ± 0.003 mm at the 95% confidence interval.

Measurement Number Output Factor Planned Shot Displacement (mm) Measured Displacement (mm) Center-to-Center Offset (mm)
1 4 mm 45.2 44.4 0.8
2 4 mm 55.9 56.6 −0.7
3 4 mm 45.1 45.9 −0.8
4 4 mm 32.0 32.5 −0.5
5 4 mm 32.0 31.5 0.5
6 8 mm 34.9 35.6 −0.7
7 8 mm 39.8 39.0 0.8