Table 4.
Planned and measured shot displacements and offsets for all seven experiments. The uncertainty in the measured center-to-center displacement was simulated using Monte Carlo techniques to be ± 0.003 mm at the 95% confidence interval.
| Measurement Number | Output Factor | Planned Shot Displacement (mm) | Measured Displacement (mm) | Center-to-Center Offset (mm) |
|---|---|---|---|---|
| 1 | 4 mm | 45.2 | 44.4 | 0.8 |
| 2 | 4 mm | 55.9 | 56.6 | −0.7 |
| 3 | 4 mm | 45.1 | 45.9 | −0.8 |
| 4 | 4 mm | 32.0 | 32.5 | −0.5 |
| 5 | 4 mm | 32.0 | 31.5 | 0.5 |
| 6 | 8 mm | 34.9 | 35.6 | −0.7 |
| 7 | 8 mm | 39.8 | 39.0 | 0.8 |