Skip to main content
. 2015 Jan 7;5:7662. doi: 10.1038/srep07662

Table 3. Comparison between two subgroups according to the right ventricle pacing sites in the patients implanted with active-fixation RV electrodes.

      Threshold (V)     Impedance (Ω)
Follow-up time n RVOT n RVA n RVOT n RVA
Baseline 43 0.82 ± 0.29 57 0.88 ± 0.33 43 604.88 ± 94.65 57 596.67 ± 94.55
1 Month 40 0.85 ± 0.66 54 0.59 ± 0.50 40 523.33 ± 136.43 54 571.08 ± 135.04
3 Months 41 1.02 ± 1.09 54 0.52 ± 0.21 41 556.10 ± 127.99 54 562.71 ± 136.16
6 Months 39 0.90 ± 1.05 57 0.62 ± 0.16 39 534.61 ± 134.87 57 549.63 ± 137.90
1 Year 42 0.95 ± 0.57 56 0.82 ± 0.19 42 504.52 ± 153.93 56 533.30 ± 140.08
2 Years 41 0.78 ± 0.24 54 0.67 ± 0.22 41 460.43 ± 135.96 54 554.70 ± 134.93
3 Years 43 0.90 ± 0.28 53 0.75 ± 0.20 43 498.07 ± 157.51 53 499.93 ± 124.80
4 Years 42 0.70 ± 0.21 56 0.65 ± 0.20 42 464.20 ± 193.31 56 544.77 ± 110.57
5 Years 40 0.65 ± 0.14 54 0.70 ± 0.22 40 532.50 ± 166.17 54 548.81 ± 105.13

Note: Data are shown as mean ± SD. RV = right ventricle; RVOT = right ventricular outflow tract; RVA = right ventricle apex.