TABLE 3.
Sensitivity and specificity for the subset of tests performed by USPHLs in B. pertussis, B. parapertussis, B. holmesii, and B. pertussis/B. holmesii (undifferentiated)
| Organism | Sample bacterial concn (CFU/ml) | na | Sensitivity (% [95% CI]) | Specificity (% [95% CI]) |
|---|---|---|---|---|
| B. pertussis | All | 907 | 96.1 (93.2–97.9) | 95.1 (93.2–96.8) |
| 1 × 103 | 170 | 88.4 (78.4–94.9) | 89.1 (81.3–94.4) | |
| 1 × 104 | 154 | 97.6 (87.1–99.9) | 99.1 (95.2–100) | |
| 1 × 105 | 312 | 97.4 (92.7–99.5) | 96.9 (93.4–98.9) | |
| 1 × 106 | 230 | 100 (90.3–100) | 94.3 (90.1–97.1) | |
| 1 × 107 | 41 | 100 (91.4–100) | —b | |
| B. parapertussis | All | 844 | 98.6 (95.9–99.7) | 99.7 (98.9–100) |
| 1 × 103 | 170 | 97.1 (84.7–99.9) | 99.3 (96.0–100) | |
| 1 × 104 | 141 | 99.0 (94.8–100) | 100 (90.3–100) | |
| 1 × 105 | 284 | — | 99.6 (98.1–100) | |
| 1 × 106 | 213 | 98.6 (92.7–100) | 100 (97.4–100) | |
| 1 × 107 | 36 | — | 100 (90.3–100) | |
| B. holmesii | All | 666 | 98.2 (94.7–99.6) | 97.8 (96.1–98.9) |
| 1 × 103 | 130 | 96.2 (86.8–99.5) | 98.7 (93.1–100) | |
| 1 × 104 | 111 | — | 96.4 (91.0–99.0) | |
| 1 × 105 | 227 | 100 (88.1–100) | 97.5 (94.2–99.2) | |
| 1 × 106 | 168 | 98.8 (93.5–100) | 98.8 (93.6–100) | |
| 1 × 107 | 30 | — | 100 (88.4–100) | |
| B. pertussis/B. holmesii | All | 330 | 97.6 (94.4–99.2) | 96.0 (90.9–98.7) |
| 1 × 103 | 78 | 93.7 (84.5–98.2) | 100 (78.2–100) | |
| 1 × 104 | 52 | 100 (73.5–100) | 95.0 (83.1–99.4) | |
| 1 × 105 | 105 | 100 (93.5–100) | 100 (92.9–100) | |
| 1 × 106 | 83 | 100 (94.3–100) | 85.0 (62.1–96.8) | |
| 1 × 107 | 12 | 91.7 (61.5–99.8) | — |
n, number of tests reported for each organism, including tests of positive and negative samples.
—, incalculable values due to empty cells, e.g., no false negatives with 95% CI.