. 2014 Dec 18;53(1):118–123. doi: 10.1128/JCM.02368-14

TABLE 4.

Sensitivity, specificity, and confidence interval for the subset of tests performed by USPHLs in all diagnostic categories, by DNA extraction method

Organism DNA extraction method na Sensitivity (% [95% CI]) Specificity (% [95% CI])
B. pertussis All 907 96.1 (93.2–97.9) 95.2 (93.2–96.8)
Automated, EasyMAG 062 100 (83.2–100) 100 (91.6–100)
Automated, MagNA Pure 371 94.3 (88.6–97.7) 95.6 (92.2–97.8)
Crude boil, heat lysis 84 100 (87.7–100) 100 (93.6–100)
Spin column, Qiagen 270 94.6 (87.9–98.2) 94.9 (90.6–97.6)
Not specified 120 100 (91.2–100) 88.8 (79.7–94.7)
B. parapertussis All 844 98.6 (95.9–99.7) 99.7 (98.9–100)
Automated, EasyMAG 77 100 (83.9–100) 100 (93.6–100)
Automated, MagNA Pure 348 98.8 (93.8–100) 100.0 (98.6–100)
Crude boil, heat lysis 84 95.2 (76.2–99.9) 98.4 (91.5–100)
Spin column, Qiagen 286 98.6 (92.5–100) 99.5 (97.4–100)
Not specified 49 100 (73.5–100) 100 (90.5–100)
B. holmesii All 666 98.2 (94.7–99.6) 97.8 (96.1–98.9)
Automated, EasyMAG 60 100 (78.2–100) 93.3 (81.7–98.6)
Automated, MagNA Pure 311 97.4 (91.0–99.7) 99.6 (97.6–100)
Crude boil, heat lysis 72 94.4 (72.7–99.9) 96.3 (87.3–99.5)
Spin column, Qiagen 175 100 (91.4–100) 97.0 (92.5–99.2)
Not specified 48 100 (73.5–100) 97.2 (85.5–99.9)
B. pertussis/B. holmesii (undifferentiated) All 330 97.6 (94.4–99.2) 96.0 (90.9–98.7)
Automated, EasyMAG 19 100 (76.8–100) 100 (47.8–100)
Automated, MagNA Pure 144 94.4 (87.4–98.2) 90.9 (80.0–97.0)
Crude boil, heat lysis 72 100 (91.6–100) 100 (88.4–100)
Spin column, Qiagen 92 100 (93.7–100) 100 (90.0–100)
Not specified 3 100 (29.2–100)
a

n, number of tests reported for each organism, including tests of positive and negative samples.