Table 2.
Reasons for aseptic revision
| Reason for revision | Number of hips (n = 131) |
|---|---|
| Aseptic loosening | 48 (37%) |
| Corrosion at a modular junction* | 40 (31%) |
| Problems related to MoM bearing | 23 (18%) |
| Retroverted acetabular component with soft tissue impingement | 5 (4%) |
| Liner dissociation | 4 (3%) |
| Instability | 4 (3%) |
| Full-thickness bearing surface wear | 3 (2%) |
| Fractured ceramic liner | 1 (0.8%) |
| Stem fracture | 1 (0.8%) |
| Leg length discrepancy | 1 (0.8%) |
| Removal of painful cable | 1 (0.8%) |
* Includes both MoM (15) and MoP (25) bearing surfaces; MoM = metal-on-metal; MoP = metal-on-polyethylene.