Skip to main content
. 2014 Dec 14;29(2):191–204. doi: 10.1002/rcm.7095

Table 1.

Typical laser parameters for this experiment. Because of the lower pulse energies available to drive the 780 nm and 776 nm transitions in Rb, these beams were focused to a smaller diameter. Beam diameters are approximate because the beam profiles were irregularly shaped. Typical pulse durations were ∼5 ns for the ablation laser and 10–15 ns for the resonance lasers

Wavelength (nm) Pulse energy (μJ) Beam diameter (mm)
Ablation
 Big Sky Centurion Nd:YAG, 5th harmonic 213 225 0.1
Sr resonance ionization
 OPOTek Opolette 355 461 1350 4
 Coherent ScanMate Pro, Fluorescein 548 dye 554 65 4
 Quantel YG-980 Nd:YAG 1064 2120 10
Rb resonance ionization
 OPPTek Opolette 355 780 3 0.5
 Contiuum ND6000, LDS-765 dye 776 2.5 0.5
 Continuum Powerlight 8020 Nd:YAG 1064 2140 10