TABLE 2.

Breakdown and final resolution of 38 discrepant results between Alere i influenza A&B test and Xpert flu A/B assaya

Test resultb
Resolution Final Alere resultb,c
Alere (38) Xpert xTAG
Flu A (3) Flu B (3) Flu B (3) Positive Flu B FP Flu A (3)
Flu B (2) Flu A (2) Flu A (2) Positive Flu A FP Flu B (2)
Flu A&B (17) Flu A (11) Flu A (11) Positive Flu A FP Flu B (11)
Flu B (6) Flu B (6) Positive Flu B FP Flu A (6)
Negative (16) Flu A (12) Flu A (12) Positive Flu A FN Flu A (12)
Flu B (4) Flu B (4) Positive Flu B FN Flu B (4)
a

The xTAG RVP was used to resolve discrepant Alere and Xpert results.

b

The test result is the type (influenza A or B) of influenza virus found if the test result was positive or negative on any of the 3 molecular assays. The number of samples with the indicated test result is shown in parentheses.

c

FP, false-positive; FN, false-negative.