Skip to main content
. 2015 Jan 23;53(2):618–625. doi: 10.1128/JCM.02443-14

TABLE 1.

Reproducibility of real-time PCR assay results

Comparison CV (%) with DNA copy no.a of:
10 102 103 104 105 106 107
Within assayb 0.9 0.5 0.2 0.4 1.2 1.5 0.5
Between assaysc 0.3 2.4 1.1 1.6 2.2 2.3 2.9
a

Tenfold dilutions of linearized plasmid DNA were used, numbers of copies per reaction mixture are shown, dilution series were thawed on 3 different days, and assays were performed in triplicate for each dilution.

b

Determined from three replicates within each assay.

c

Determined from three independent assays performed on different days.