Table 7.

Accordance of Cu(II), Cd(II), Pb(II) and Zn(II) ions spiked in Jeram landfill leachate by both developed A. fischeri-based biosensor and AAS method.

Heavy Metal Added (μg/L) Biosensor (n = 3) AAS (n = 3) Calculated t-test

Found (μg/L) Accordance (%) Found (μg/L) Accordance (%)
Cu(II) 0 11.59 ± 1.16 - 11.96 ± 0.96 - 0.307
20 29.32 ± 2.73 88.65 30.43 ± 2.47 92.35 0.776
30 36.76 ± 7.64 83.90 37.98 ± 2.61 86.73 0.280
50 63.16 ± 6.19 103.08 61.83 ± 4.88 99.74 0.468
70 85.50 ± 9.95 105.59 84.45 ± 6.11 103.56 0.328
90 99.45 ± 13.03 97.62 97.03 ± 7.96 94.52 0.937
100 108.39 ± 15.24 96.80 106.85 ± 8.04 94.89 0.452

Cd(II) 0 214.21 ± 3.30 - 215.103 ± 3.55 - 0.241
200 411.35 ± 24.25 98.86 409.50 ± 14.45 97.20 0.388
300 488.97 ± 33.28 91.57 484.47 ± 23.46 89.57 0.921
500 641.49 ± 51.47 85.45 650.94 ± 42.52 87.17 0.238
700 888.01 ± 68.54 96.26 879.60 ± 47.84 94.93 0.741
900 1077.01 ± 95.22 95.86 1080.03 ± 47.89 96.10 0.023
1000 1266.27± 129.12 105.21 1259.91 ± 83.81 104.58 0.069

Pb(II) 0 54.72 ± 1.44 - 55.19 ± 1.36 - 0.535
60 104.89 ± 11.21 83.60 106.36 ± 4.41 85.28 0.217
70 116.82 ± 13.37 88.71 118.22 ± 5.87 90.04 0.175
80 138.22 ± 17.15 104.36 137.03 ± 6.51 102.30 0.137
100 166.12 ± 21.95 111.39 163.48 ± 9.58 108.28 0.349
200 251.55 ± 23.39 98.41 247.69 ± 16.52 96.25 0.262
500 474.67 ± 67.12 94.93 532.43 ± 35.91 95.45 0.066

Zn(II) 0 64.28 ± 1.43 - 63.71 ± 0.87 - 0.479
50 110.42 ± 6.71 92.28 110.72 ± 4.27 94.02 0.283
70 131.78 ± 10.26 96.41 132.44 ± 5.51 98.18 0.262
80 132.73 ± 13.65 85.56 133.45 ± 6.17 87.17 0.183
90 142.02 ± 15.37 86.37 143.32 ± 7.15 88.45 0.192
100 155.43 ± 18.25 91.15 157.62 ± 8.83 93.90 0.499
200 274.13 ± 26.29 104.92 267.60 ±17.03 101.95 1.077

Notes: The critical value, t4 = 2.78 (P = 0.05, 95%). The linear equation of [Cu(II)]biosensor versus [Cu(II)AAS, [Cd(II)]biosensor versus Cd(II)AAS, [Pb(II)]biosensor versus [Pb(II)]AAS and [Zn(II)]biosensor versus [Zn(II)]AAS were [Cu] biosensor = 0.9589[Cu]AAS − 1.3885, [Cd]biosensor = 0.9965[Cd]AAS − 0.1318, [Pb]biosensor = 1.0057[Pb]AAS − 1.937 and [Zn]biosensor = 0.9546[Zn]AAS − 4.4235, respectively. The R2 values for Cu(II), Cd(II), Pb(II) and Zn(II) were 0.9996, 0.9996, 0.9998 and 0.9991, respectively.