Skip to main content
. 2012 Dec 1;6:49–54. doi: 10.1016/j.ijsp.2011.11.002

Table 4.

Results of independent review of 3D scans for assessment of position of 602 pedicle screws

Grade No. of screws (%) Intraoperative action performed
Optimal (no breach) 559 (92.9) None revised
Acceptable (breach ≤2 mm) 32 (5.3) 6 of 32 (19%) revised to optimal position
Unacceptable (breach >2 mm) 11 (1.8) All revised to optimal/acceptable position
Total 602 (100)