Table 2.

MPO and reference thresholds for implantable BCDs

Device MPO in force level [dB re 1μN]
Comment/reference
0.5 k 1 k 2 k 4 k PTA4
1. Bonebridge MPO @ Skullsimulator 93 104 102* 84 95.8 Håkansson B, personal communication, 2014
2. BCI MPO @ Skullsimulator 97 102 94* 95 97.0 Taghavi et al, unpublished data, 2014
3. RETFL DBC @ pos A (BAHA) 48 45.5 26 27.5 36.8 Carlsson and Håkansson79
4. Correction: Pos A (BAHA) vs pos B (BCI) [dB] 3 9 10 8 7.5 Reinfeldt et al50
5. RETFL DBC @ pos B (BCI) 45 36.5 16 19.5 29.3 Calculated (3 minus 4)

Notes: Pos A refers to the approximate skull position for a BAHA, 55 mm behind and slightly above the ear-canal opening. Pos B is the skull position for the BCI, 10–15 mm behind the ear-canal opening. Correction (4) is the calculation of the sensitivity difference between stimulation of the BCI and the BAHA positions (pos B and pos A, respectively). @, attached to.

*

Values were taken slightly below 2 kHz to avoid interference with a resonance peak.

Abbreviations: BAHA, bone-anchored hearing aid; BCI, bone-conduction implant; BCD, bone-conduction device; DBC, direct bone conduction; MPO, maximum power output; pos A, position A; pos B, position B; PTA4, four-frequency pure-tone average; re, relative to; RETFL, reference equivalent threshold force level.