Table 8.

Ventilation parameters, airway parameters, and Brimacombe scores of the devices.

Ventilation parameters
Parameter ProSeal LMA I-gel
Neutral Flexion Extension Lateral rotation Neutral Flexion Extension Lateral rotation
Delivered tidal volumes (mL) 461 ± 68.72 452 ± 100.75 457 ± 66.28 463 ± 66.26 481 ± 48.69 481 ± 52.67 477 ± 50.69 478 ± 49.63
Peak airway pressures (cm H2O) 16 ± 2.52 19 ± 6.09 15 ± 2.85 16 ± 2.59 15 ± 2.99 17 ± 5.25 15 ± 3.39 16 ± 3.24
EtCO2 (mm Hg) 31 ± 2.33 31 ± 1.98 31 ± 2.24 31 ± 2.06 31 ± 2.27 31 ± 2.02 31 ± 2.10 31 ± 2.07
LAW-PAW (cm H2O) 11 ± 5.02 12 ± 6.83 9 ± 5.18 12 ± 4.27 6 ± 4.86* 7 ± 6.23* 3 ± 4.46* 6 ± 4.77*
Ventilation score (3/2/1/0) 30/0/0/0 29/0/0/1 30/0/0/0 30/0/0/0 30/0/0/0 29/1/0/0 30/0/0/0 30/0/0/0

Data shown are mean ± SD or numbers. P < 0.05 is considered significant. * P value < 0.01.

(LAW-PAW, compared with similar positions between I-gel and ProSeal LMA.) LAW-PAW was consistently better for ProSeal LMA than I-gel in all positions. Other parameters including peak airway pressures, tidal volume delivery, and ventilation scores were comparable between the two groups in all positions.