Skip to main content
. Author manuscript; available in PMC: 2016 Feb 15.
Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2015 Jan 2;0:19–26. doi: 10.1016/j.jchromb.2014.12.021

Table 2.

Intra-day and inter-day accuracy and precision of EHop-016 in mouse plasma.

Analyte Spiked
Level
(ng/mL)
Intra-Day (n=6) Inter-Day (n=12)

Accuracy
(RE %)
Precision
(RSD %)
Accuracy
(RE %)
Precision
(RSD %)
5 7.1 3.5 0.7 2.4
10 1.4 2.0 −2.9 5.5
EHop-016 30 4.0 1.8 3.1 1.7
75 0.2 0.7 2.3 3.4
750 5.2 1.9 9.5 3.1
1000 6.2 1.0 −0.3 5.2