. 2015 Feb;56(2):403–412. doi: 10.1194/jlr.M055608

TABLE 2.

Assay validation parameters for 27OHC3S

Nanograms per Milliliter (Spiked Concentration) Mean (Concentration Measured) SD Precision (% CV) Accuracy (% RE)
Intra-assay (n = 5) 5 4.9 0.6 11.3 −1.4
50 52.8 2.2 4.2 5.6
Inter-assay (n = 12) 5 4.9 0.7 14.8 −2.4
50 52.3 3.4 6.4 4.5
Percent Recovery 5 94.2
50 100.6
Linearity, LOQs, LOD (ng/ml) Linearity LOD LLOQ ULOQ
2.5–500 1 2.5 500
Matrix effect study (regression line) y = 0.975x + 0.018 (R2 = 0.999) (2.5–500 ng/ml)

CV, Coefficient of variation; RE,: relative error; LLOQ, lower limit of quantification; ULOQ, upper limit of quantification.